Reference Oscillator Frequency

[:SENSe]:OPTion:ROSCillator:EXTernal:FREQuency <Hz>
[:SENSe]:OPTion:ROSCillator:OUTPut?
Set the frequency of the external reference oscillator.
Preset
and *RST:
10 MHz
Maximum Value: 30 MHz
Minimum Value: 1 MHz
Front Panel
Access:
Mode Setup, Demod...
or
Det/Demod, More

Reference Oscillator Rear Panel Output

[:SENSe]:OPTion:ROSCillator:OUTPut[:STATe] OFF|ON|0|1
[:SENSe]:OPTion:ROSCillator:OUTPut?
This enables or disables the 10 MHz output reference signal from the
rear panel port labelled
connected to the
10 MHz REF IN
reference oscillator source is set to external mode.
Preset
and *RST:
Off
Front Panel
Access:
Mode Setup, Demod...
or
Det/Demod, More

Reference Oscillator Source

[:SENSe]:OPTion:ROSCillator:SOURce INTernal|EXTernal
[:SENSe]:OPTion:ROSCillator:SOURce?
Select the reference oscillator (time base) source to the internal
reference oscillator or an external oscillator. Use
OPT:ROSC:EXT:FREQ to tell the instrument the frequency of the
external reference.
Chapter 7
Programming Language Reference
. This output signal needs to be
10 MHz OUT
port on the rear panel whenever the
SENSe Subsystem