Figure 4-4. Selecting A Tube For Rotor And Tube Specifications - Beckman Coulter Optima XL Series Instruction Manual

Preparative ultracentrifuge
Table of Contents

Advertisement

Rotor Functions
4-8
NOTE
The full-size tube is always given as first in the list and is
not a g-Max tube.
3. When a specific tube is highlighted, the specifications for that rotor/tube
combination appear in the window on the left. Figure 4-4 shows the
specifications for the Type 90 Ti rotor with a 6.5 mL tube.
✶✪✲✷❈✼✬✻✮✮✷❛
4. Press
return to the ROTOR CATALOG window.
ROTOR SPECIFICATIONS
MODEL: Type 90Ti
MAX RPM: 90000
k FACTOR: 14
RCF AT MAX SPEED:
rMIN: 34.2 mm . . . 310,300 x g
rAVG: 55.3 mm . . . 501,700 x g
rMAX: 76.5 mm . . . 694,000 x g
NOMINAL VOLUME: 6.5 mL
TUBE COUNT: 8
ROTOR VOLUME: 52.0 mL
TUBE ANGLE: 25.0 deg
00 RPM
0:00 H:M
26.2 DEG
U S E A R R O W K E Y S T O R E V I E W R O T O R S P E C I F I C A T I O N S W I T H G - M A X T U B E O P T I O N S .

Figure 4-4. Selecting a Tube for Rotor and Tube Specifications

to return to the Main Screen or
G-MAX POLYALLOMER TUBES
WITH G-MAX TUBE
CLASS: FGHRS
QUICK SEAL
OPTISEAL
*MAXIMUM VOLUME TUBE
DOWN
UP
✻✸✽✸✻❈✬✪✽✪✵✸✰❛
to
Fill Volume
Tube Part No.
12.5 mL
342413*
6.5 mL
345830
4.7 mL
356562
8.9 mL
361623
ROTOR
MAIN
CATALOG
SCREEN

Advertisement

Table of Contents
loading

Table of Contents