7.4 Replacing the Main Power Fuses

Warning:
Avertissement: Avant de remplacer les fusibles, arrêtez l'instrument. Assurez-vous
1. Remove the fuse cartridge, using a small screwdriver.
2. Replace the fuses.
Warning:
Avertissement: Installez toujours deux nouveaux fusibles. Utilisez uniquement
Description
Fuse, 2A, slow-blow (5 x 20 mm)
3. Reinstall the fuse cartridge.
4. Reconnect the power cord to its source. Turn on the power to the TCC.
Operating Instructions
TCC-3000SD and TCC-3000RS Column Compartments
Turn off the TCC. Disconnect the power cord from its source.
de bien débrancher le cordon d'alimentation de la source secteur.
Fig. 52: Fuse cartridge
Always install two fuses. Use only the fuses indicated in the
following table.
les fusibles indiqués ci-dessous.
Fuse cartridge
Part No.
Included in Fuses Kit (part no. 6710.9001)
For information about the kit, see section 9.3
(→ page 120).
UltiMate 3000 Series:
Page 109