Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual page 97

Column compartments
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

Message
Error while programming
flash!
Flash object error!
Gas leak alarm!
Gas leak sensor error!
Humidity leak alarm!
Humidity leak sensor error!
I2C bus error!
(This message may be
extended by additional text.)
I2C bus device x timeout!
(where x = chip card,
temperature sensor, or valve)
Leak temperature sensor
communication error!
Left valve controller no
communication.
Left valve controller
configuration error!
Left valve moving timeout
error!
Rear temperature sensor
communication error!
Operating Instructions
TCC-3000SD and TCC-3000RS Column Compartments
Remedial Action
(This message may be extended by additional text.)
Turn the TCC off and on again by pressing the power switch on the
rear of the TCC and retry.
(This message may be extended by additional text.)
Turn the TCC off and on again by pressing the power switch on the
rear of the TCC and retry.
The gas sensor detected excessive gas in the column chamber.
Eliminate the cause. Tighten leaking connection and vent the column
chamber.
The gas sensor may be influenced by the drain tubing from other
modules in the UltiMate 3000 system. Verify that the drain tubing of
the TCC is connected as described in the installation instructions
provided the system drain kit. As an alternative, connect separate
drain tubing to both the TCC and the system module stacked upon
the TCC.
Turn the TCC off and on again by pressing the power switch on the
rear of the TCC and retry.
The humidity sensor reports an increased concentration of humidity
in the column chamber. Eliminate the cause. Tighten leaking
connection and vent the column chamber (→ page 22).
Turn the TCC off and on again by pressing the power switch on the
rear of the TCC and retry.
Turn the TCC off and on again by pressing the power switch on the
rear of the TCC and retry.
Turn the TCC off and on again by pressing the power switch on the
rear of the TCC and retry.
Turn the TCC off and on again by pressing the power switch on the
rear of the TCC and retry.
Turn the TCC off and on again by pressing the power switch on the
rear of the TCC and retry.
Turn the TCC off and on again by pressing the power switch on the
rear of the TCC and retry.
Check whether the spanner has been tightened too firmly. If the
spanner is too tight, the valve cannot be switched. Loosen the
spanner a bit and retry.
Turn the TCC off and on again by pressing the power switch on the
rear of the TCC and retry.
Turn the TCC off and on again by pressing the power switch on the
rear of the TCC and retry.