Installing A Separation Column - Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual

Column compartments
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

UltiMate 3000 Series:
TCC-3000SD and TCC-3000RS Column Compartments

4.6 Installing a Separation Column

Warning:
Avertissement: Ne touchez à aucune partie en métal ou plastique à l'intérieur du
Four column brackets are installed in the column chamber at the factory. Attach the
columns to the brackets using the special column clips from the TCC accessories kit.
1. The column clip consists of two pieces. To remove the ring from the bottom part,
slightly press the ring in the direction of the arrows.
2. Attach the bottom part of the clip to the bracket and turn it 90° clockwise.
Page 50
Before carrying out any work in the column chamber, wait for the
column chamber to cool down. Do not touch any metal or plastic
parts inside the column chamber while the temperature is higher than
50 °C.
compartiment de colonne tant que le point de réglage de température
est > 50 °C. Attendez que le four refroidisse, par exemple, avant de
changer une colonne ou avant de procéder à tous travaux de
maintenance.
Fig. 18: Two-piece column clip
Fig. 19: Installing the bottom part
Press the ring in the direction of the arrows. (The
arrows on the ring are printed in black.)
Operating Instructions