4. On the General page (→ page 38), click Firmware Download. A message displays
the firmware version that is currently installed in the TCC and the version that will be
downloaded from Chromeleon.
Tip: If the TCC comes with a newer firmware than the version included in
Chromeleon, do not downgrade the firmware. Older firmware may be
incompatible with new hardware revisions.
5. Click Yes to start the download. (Click No to cancel the action.)
The download may take several minutes. The download is complete when Download
finished successfully appears in the Messages Server window in the Chromeleon Server
Configuration program. The message appears also in the Chromeleon Audit Trail.
If the download from Chromeleon is not successful, the related messages appear in the Audit
Trail. In this case, turn the TCC off and on. Repeat the download as described above. If the
download fails again, contact Thermo Fisher Scientific Service for Dionex HPLC Products.