5.4.2.4 Information Menu
The Information menu provides information for diagnostic purposes (read-only).
To ...
See general information about the TCC, for example,
firmware version, serial number, operating hours, and the last 10 errors.
See general information about the switching valves, for example,
configuration and switching cycle counter.
See information about the temperature, for example
working load and cooling load.
5.4.2.5 Configuration Menu
The Configuration menu provides information about the configuration of the TCC and
allows you to make the required settings or change the settings.
To ...
Restore the factory settings. In the Reset to factory defaults? dialog box, select OK
to confirm the restore or select Cancel to keep your settings.
Set the display and function key parameters:
Brightness—sets the screen brightness. (The input is in per cent.)
Contrast—sets the screen contrast. (The input is in percent.
Key sound—sets whether a beep sounds when you select a function key.
Key autorepeat—sets whether the keystroke is automatically repeated when you
remain on the key for a longer period, for example, to change a value quickly (On =
yes or Off = no).
Valve display—sets the appearance of the status screen. For the status screen shown
in Fig. 31 (→ page 60), select Big. For the screen shown in Fig. 42 (→ page 78),
select Small.
Specify which valve is present in the right and/or left position. The entry consists the
valve type (= HT for TitanHT valves or HP for TitanHP valves) and the number of
valve ports and positions.
Examples:
If a TitanHT 2-position, 10-port valve is installed, select HT 10-port, 2-pos.
If a TitanHP 2-position, 6-port valve is installed, select HP 6-port, 2-pos.
If no valve is installed, select Not installed.
Operating Instructions
TCC-3000SD and TCC-3000RS Column Compartments
UltiMate 3000 Series:
Select ...
System
Valve
Temperature work load
Select ...
Reset to factory
defaults
Display & soft
keys
Valve
Page 71