UltiMate 3000 Series:
TCC-3000SD and TCC-3000RS Column Compartments
7.3.2 Replacing the Valve Pod
Valve pod* for
TitanHP™ column switching valves
suitable for pressures < 41 MPa (6000 psi)
2-position, 6-port valve (SST)*
2-position, 10-port valve (SST)*
6-position, 7-port valve (Ti)
The rotor seal in the valve pod is suitable for pH values < 10.
* To use these valves for applications with pH values > 10, replace this rotor seal
with the PEEK rotor seal and stator face seal (→ page 107).
Biocompatible TitanHP™ column switching valves (PEEK)
suitable for pressures < 34 MPa (5000 psi)
2-position, 6-port valve (PEEK)
2-position, 10-port valve (PEEK)
The rotor seal in the valve pod is suitable for pH values 0 - 14.
TitanHT™ column switching valves
suitable for pressures < 103 MPa (15000 psi)
2-position, 6-port valve (SST)
2-position, 10-port valve (SST)
6-position, 7-port valve (SST)
The rotor seal in the valve pod is suitable for pH values 0 - 14.
Biocompatible TitanHT™ column switching valves,
suitable for pressures < 103 MPa (15000 psi)
2-position, 6-port valve (Ti)
2-position, 10-port valve (Ti)
6-position, 7-port valve (Ti)
The rotor seal in the valve pod is suitable for pH values 0 - 14.
TitanHT™ column switching valve,
suitable for pressures < 125 MPa (18130 psi)
6-position, 7-port valve (SST)
The rotor seal in the valve pod is suitable for pH values 0 - 10.
*
All valve pods are shipped with appropriate fitting connections.
Page 100
Part No.
6722.9013
6722.9023
6722.9035
6723.9013
6723.9023
6730.0006
6730.0026
6730.0016
6730.0031
6730.0032
6730.0030
6730.0050
Operating Instructions