Connections On A 2-Position, 6-Port Column Switching Valve - Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual

Column compartments
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

UltiMate 3000 Series:
TCC-3000SD and TCC-3000RS Column Compartments
To ensure proper connection, a short part of the capillary must stick out of the ferrule.
Therefore, remove the capillary from the valve to verify this. In addition, check that the
ferrule grabbed the capillary. When pulling the ferrule gently, you should not be able to
move the ferrule on the capillary.
3. Route the capillaries through the capillary guides that are provided on the left and right
of the column chamber (→ Fig. 3, page 21) into the column chamber. The connections
should be as short as possible.
4. Reinstall the drip tray of the valve if applicable.

4.8.1 Connections on a 2-position, 6-port Column Switching Valve

Fig. 26 shows a typical example for the connections on a 2-position, 6-port column
switching valve.
Fig. 26: Example for the connections on a 2-position, 6-port column switching valve
If you have connected the inlet capillary of a column to position 6 of a 2-position, 6-port
switching valve, you must use position 5 for the outlet capillary of this column. The same
applies to valve ports 2 and 3.
Page 54
Fig. 25: Capillary with fitting screw and ferrule
for connection to the biocompatible valve pod
Port
Connect the capillary ...
1
from the autosampler or pump
2
to column B or
to the pre-column heater
3
from the outlet of column B
4
to the post-column cooler or
to the detector
5
from the outlet of column A
6
to column A or
to the pre-column heater
Operating Instructions

Advertisement

Table of Contents
loading

This manual is also suitable for:

Tcc-3000sdTcc-3000rs

Table of Contents