4.8 Connecting the Components to the Column Switching Valve

Warning:
Before carrying out any work in the column chamber, wait for the
column chamber to cool down. Do not touch any metal or plastic parts
inside the column chamber while the temperature is higher than 50 °C.
Avertissement: Ne touchez à aucune partie en métal ou plastique à l'intérieur du
1. For easier access to the connection ports on the switching valve, you may unhook the
drip tray of the valve.
2. Connect the capillaries to the column switching valve, by using an appropriate tool.
For example configurations for the different valve types, see sections 4.8.1 and 4.8.3.
When connecting the capillaries, also observe the general precautions on page 46.
Only biocompatible column switching valve (TitanHP, PEEK)
To connect the capillaries to the biocompatible column switching valve, install only
PEEK fitting connections and/or the ferrules provided in the accessories kit for the valve
or original Dionex spare parts. Do not use metal fittings and ferrules for the connection.
They can cause damage to the valve.
a) Slide the fitting nut and ferrule onto the capillary.
b) Insert the assembly into the valve port.
c) Push the capillary all the way into the port, and then tighten the connection until the
ferrule grabs the capillary.
Operating Instructions
TCC-3000SD and TCC-3000RS Column Compartments
compartiment de colonne tant que le point de réglage de température
est > 50 °C. Attendez que le four refroidisse, par exemple, avant de
changer une colonne ou avant de procéder à tous travaux de
maintenance.
Fig. 24: Unhooking the drip tray