5.5.5 Adjusting the Sensitivity of the Gas and Humidity Sensors

You can adjust the sensitivity of the gas or humidity sensor in Chromeleon or on the front
panel display.
To adjust the sensitivity from Chromeleon
1. In Chromeleon, open the Commands dialog box for the TCC.
2. Select GasLeakSensor or select HumidityLeakSensor and set the sensor sensitivity:
Select Low, Standard, or High. When the sensor is activated
♦ The Status LED on the front panel door is red.
♦ A message appears in Chromeleon and on the TCC display.
♦ A beep sounds.
Select Low_Silent, Standard_Silent, or High_Silent. When the sensor is activated
♦ The Status LED on the front panel door is red.
♦ A message appears in Chromeleon and on the TCC display.
♦ No beep sounds.
Select Off to deactivate leak detection.
To adjust the sensitivity on the front panel display
1. Show the function keys and select Leak.
2. Select Gas leak mode or Humidity leak mode.
3. Select low, standard, or high, or select low silent, standard silent, or high silent.
(For details about these settings, see further up in this section.) To deactivate leak
detection, select off.