Replacing The Main Power Fuses - Thermo Scientific Corona Veo Operating Manual

Charged aerosol detectors
Table of Contents

Advertisement

7
Maintenance and Service

7.10 Replacing the Main Power Fuses

When
Parts required
Tools required
Preparations
Page 196
Blown fuses
Replacement fuses (2 fuses, slow-blow, 1 A, 5 x 20 mm) from Fuses Kit
Slotted screwdriver, any size between 3.3 mm and 5.5 mm is
appropriate
WARNING—Electric Shock
High voltages are present inside the detector that could cause an electric
shock or damage to the device.
Turn off the detector with its main power switch. Disconnect the power
cord from both the power source and the device.
Use only the fuses of the type and current rating specified for the
detector by Thermo Fisher Scientific. Do not use repaired fuses and do
not short-circuit the fuse holders.
AVERTISSEMENT—Décharge électrique
Des tensions élevées sont présentes dans le détecteur et peuvent causer
des chocs électriques ou endommager l'appareil.
Arrêtez le détecteur au moyen de son interrupteur d'alimentation
principal. Déconnectez le cordon d'alimentation de la source
d'alimentation et de l'appareil.
Utilisez uniquement les fusibles du type et de courant nominal tels que
spécifié par Thermo Fisher Scientific pour le détecteur. N'utilisez pas des
fusibles réparés et ne court-circuitez pas les porte-fusibles.
Charged Aerosol Detectors Corona Veo (RS) Operating Manual