4.8.3 Connections on a Multi-Position Column Switching Valve

Fig. 28 shows the connections on a multi-position column switching valve (here as an
example, on a 6-position, 7-port valve).
Fig. 28: Example for the connections on a multi-position valve (here on a 6-position, 7-port valve)

4.9 Connecting Drain Tubing

To discharge liquid leaks that might have accumulated in the interior, the TCC has a drain
port at the bottom right of the instrument.
Direct liquid leaks to waste through the drain system of the UltiMate 3000 system, by
using the components from the drain kit. The kit is shipped with the UltiMate 3000 pumps
and can be ordered separately (part no. 6040.0005). The kit includes all required
components and detailed installation instructions.
Page 56
Port
Connect ...
Column 1
1
Column 2
2
Column 3
3
Column 4
4
Column 5
5
Column 6
6
Valve inlet capillary, or valve outlet capillary or
7
capillary to the post-column cooler insert
Fig. 29: Drain port