Column compartments
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

UltiMate 3000 Series:
TCC-3000SD and TCC-3000RS Column Compartments
Problem
Impaired performance of the
post-column cooler
Page 94
Probable Cause
The post-column cooler insert is
not installed correctly.
The post-column cooler insert is
blocked or defective.
The opening for the post-column
cooler insert is defective.
Remedial Action
Verify that the post-column
cooler is installed correctly.
The label must face away
from the instrument
(→ page 52).
Install a new post-column
cooler insert.
Contact Service.