Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual page 50

Column compartments
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

UltiMate 3000 Series:
TCC-3000SD and TCC-3000RS Column Compartments
Configuration Page
Use this page to determine whether Chromeleon shall record the temperature signals as
separate channels and select the pressure unit.
Temperature Signal
The check box is selected by default. Accept this setting if you want to record the
temperature of the TCC as a separate channel (→ page 83).
The name under which temperature signal is identified in the installation environment
and in the Chromeleon client is ColumnOven_Temp. Accept the default name. If you
enter a different name, you may have to re-link the controls on the control panels.
Cooler Temp. Signal (only TCC-3000RS)
The check box is selected by default. Accept this setting if you want to record the
temperature of the post-column cooler as a separate channel (→ page 83). The signal will
be recorded only if you have also selected the post-column cooler check box on the
Components page (→ page 39). Accept the default name (Cooler_Temp) under which
this temperature signal is identified in the installation environment and in the
Chromeleon client. If you change the default name, you may have to re-link the
corresponding controls of the related control panels.
Pressure Unit
Select the pressure unit to be used for storing the pressure values on the column ID chip
card.
Rear Temperature
The check box is selected by default. Accept this setting. The signal will be recorded as a
separate channel. If a problem occurs, the channels can provide helpful information to
identify and eliminate the source for the problem.
Page 42
Fig. 12: Configuration page (here TCC-3000RS)
Operating Instructions