Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual page 98

Column compartments
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

UltiMate 3000 Series:
TCC-3000SD and TCC-3000RS Column Compartments
Message
Right valve controller no
communication.
Right valve controller
configuration error!
Right valve moving timeout
error!
When the TCC is operated from Chromeleon and if communication between Chromeleon
and the TCC cannot be established, messages appear in the Chromeleon Audit Trail.
Message
TCC-3x00@USB-1610103 -
Device not found on the USB.
Error opening TCC-3x00
@USB-1610103 – The System
cannot find the file specified
Error issuing control request to
TCC-3x00@USB-1610103
Error reading from TCC-3x00
@USB-1610103
Data error (cyclic redundancy
check)
Error reading from TCC-3x00
@USB-1610103
Page 90
Remedial Action
Turn the TCC off and on again by pressing the power switch on the
rear of the TCC and retry.
Turn the TCC off and on again by pressing the power switch on the
rear of the TCC and retry.
Check whether the spanner has been tightened too firmly. If the
spanner is too tight, the valve cannot be switched. Loosen the
spanner a bit and retry.
Turn the TCC off and on again by pressing the power switch on the
rear of the TCC and retry.
Remedial Action
The USB connection between the TCC and the Chromeleon server
may be interrupted. Check the USB connection.
The power supply to the TCC may be interrupted. Check the power
supply connection of the TCC.
The USB connection between the TCC and the Chromeleon server
may be interrupted. Check the USB connection.
The power supply to the TCC may be interrupted. Check the power
supply connection of the TCC.
The USB connection between the TCC and the Chromeleon server
may be interrupted. Check the USB connection. Check the power
supply connection of the TCC.
Remove the TCC specified in the message from the server
configuration or else, select a different TCC from the list of available
TCC in the server configuration program.
Check the USB connection. The connection to the next hub must not
exceed 5 m. The overall connection length, including the hub
connections must not exceed 30 m.
Replace any defective USB cable or hub.
The USB connection between the TCC and the Chromeleon server
may be interrupted. Check the USB connection.
The power supply to the TCC may be interrupted. Check the power
supply connection of the TCC.