UltiMate 3000 Series:
Fluorescence Detectors FLD-3100 and FLD-3400RS
7.5 Replacing the Main Power Fuses
Warning:
Avertissement: Avant de remplacer les fusibles, arrêtez le détecteur. Assurez-vous
1. Remove the fuse cartridge by using a small screwdriver.
2. Replace the fuses.
Warning:
Avertissement:
Description
Überlastsicherung,
2A, träge, 5 x 20 mm
3. Reinstall the fuse cartridge.
4. Reconnect the power cord to its source. Turn on the detector power.
Page 114
Turn off the main power switch. Disconnect the power cord from its
source.
de bien débrancher le cordon d'alimentation de la source secteur.
Fig. 41: Fuse cartridge
Always install two new fuses. Use only the fuses indicated in
the following table.
Installez toujours deux nouveaux fusibles.
Utilisez uniquement les fusibles indiqués ci-dessous.
Part no.
Included in Fuses Kit (part no. 6074.0005)
For information about the kit, see section 9.3
(→ page 121).
Fuse cartridge
Operating Instructions