Replacing The Main Power Fuses - Thermo Scientific Vanquish VC-D40 Operating Manual

Variable wavelength detectors
Table of Contents

Advertisement

7
Maintenance and Service
7.9

Replacing the Main Power Fuses

When
Parts required
Tools required
Preparations
Follow these steps
Page 122
Blown fuses
Fuses (2 fuses, 3.15 AT, 250 V AC, slow-blow, 5 x 20 mm) from Fuses Kit
Slotted screwdriver, any size between 3.3 mm and 5.5 mm is
appropriate
WARNING—Electric Shock
High voltages are present inside the device that could cause an electric
shock or damage to the device.
Turn off the device with its main power switch. Disconnect the
power cord from both the power source and the device.
Use only the fuses of the type and current rating specified for the
device by Thermo Fisher Scientific.
Do not use repaired fuses and do not short-circuit the fuse holders.
The fuse holder is located next to the main power switch.
Figure 26: Fuse holder
No.
Description
1
Main power switch (on/off control)
2
Fuse holder
3
Power-inlet connector
Variable Wavelength Detectors (VC-D40, VF-D40)
Operating Manual