Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual page 31

Column compartments
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

A 2-position, 10-port valve is used for applications that require overlapping injections
with offline re-equilibration of the column.
Injector
From column re-
equilibration
pump
Position 10-1
Fig. 5: Example for column switching with a 2-position, 10-port column switching valve
In position 10-1, column 1 analyzes while column 2 re-equilibrates. In position 1-2,
column 2 analyzes while column 1 re-equilibrates.
Two multi-position valves are the appropriate choice for applications with several
columns, for example, for method scouting.
Injector
Fig. 6: Example flow schematic for column connection on two multi-position valves
For information about how to connect the capillaries to the column switching valves, see
section 4.2 (→ page 46).
Operating Instructions
TCC-3000SD and TCC-3000RS Column Compartments
Column 2
Detector
Column 1
Column 6
Column 1
Column 5
Column 4
Column 3
(here 6-position, 7-port valves)
Waste
Injector
From column re-
equilibration
pump
Position 1-2
Column 2
UltiMate 3000 Series:
Column 2
Waste
Detector
Column 1
Detector
Page 23