5.9

Replacing the Main Power Fuses

HIGH VOLTAGE—Disconnect the main power cord from its source and
also from the rear panel of the Dionex PDA.
HAUTE TENSION—Débranchez le cordon d'alimentation principal de
sa source et du panneau arrière du Dionex PDA.
HOCHSPANNUNG—Ziehen Sie das Netzkabel aus der Steckdose und
der Netzbuchse auf der Rückseite des Dionex PDA.
1. Press the power button on the front of the Dionex PDA to turn off the power.
2. Turn off the main power switch on the Dionex PDA rear panel.
3. Disconnect the power cord from the Dionex PDA rear panel.
4. The fuse holder is part of the main power receptacle (see
recessed lock on each side of the fuse holder.
Using a small screwdriver (or your fingernails), push one of the locks toward
the center to release it. (The fuse holder will pop out slightly when the lock is
released.) Repeat for the other lock. Pull the fuse holder straight out of the
compartment.
Locking
Spring
Insert
screwdriver
and twist to
release
(each side)
Doc. 065378-01 1/12
Fuse Holder
Key
Main Power
Receptacle
Figure 5-4. Main Power Fuse Holder
Figure
Locking
Fuses (2)
Spring
Fuse Holder
(Side View)
5 • Service
5-4). Note the
103