3.4 Connecting the TCC

3.4.1 General Information

If you want to operate the TCC from Chromeleon
Before you connect the TCC to the USB port on the Chromeleon computer and turn on the
TCC power, verify that Chromeleon is installed on the computer and that the license code is
entered. Only if you install Chromeleon first, the USB driver for the TCC is automatically
loaded and the Windows

3.4.2 Connecting the USB Cable

Connect the TCC to the Chromeleon computer via the USB ports on the rear panel (→ Fig.
2, page 19). Select one of the following alternatives:
Connect the TCC directly to the USB port on the Chromeleon computer.
Connect the TCC to the internal USB port on another module in the UltiMate 3000
system that is connected to the Chromeleon computer (→ page 32).
To ensure trouble-free operation, use only these cables for the connection (the cables are
provided in the accessories kit for the TCC):
USB Cable
USB cable, type A to type B, high speed USB 2.0 (cable length: 1 m)
USB cable, type A to type B, high speed USB 2.0 (cable length: 0.5 m)
Tip: Each USB device can be separated from the computer or next USB hub by no
more than 5 meters.
Operating Instructions
TCC-3000SD and TCC-3000RS Column Compartments
®
operating system can detect the TCC when the power is turned on.