Turning On Post-Column Thermostatting - Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual

Column compartments
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

UltiMate 3000 Series:
TCC-3000SD and TCC-3000RS Column Compartments

5.5.2 Turning on Post-Column Thermostatting

Only TCC-3000RS
You can turn post-column thermostatting on and off and set the temperature in Chromeleon or
on the front panel display
To turn on post-column thermostatting from Chromeleon
1. On the Components page in the Properties dialog of the TCC, verify that the Post-
column Cooler check box is selected (→ page 39).
2. In Chromeleon, open the Commands dialog box for the TCC.
3. Select ColumnOven.
4. Select Cooler_Temperature and Nominal, and then enter the temperature.
Cooler_TempCtrl is automatically set to On. Post-column thermostatting is active.
Set Cooler_TempCtrl to Off if you do not want to use post-column thermostatting for a
certain application.
If you want to use post-column thermostatting later again, set Cooler_TempCtrl to On.
When you change the temperature setting under Nominal, Chromeleon sets
Cooler_TempCtrl automatically to On.
To turn on post-column thermostatting on the front panel display
1. Show the function keys and select Temp.
2. Select Post-cooler and enter the temperature. Post-cooler control is automatically set to
On. Post-column thermostatting is active.
Set Post-cooler control to Off if you do not want to use post-column thermostatting for a
certain application.
If you want to use post-column thermostatting later again, set Post-cooler control to On.
When you change the temperature setting under Post-cooler, Post-cooler control is
automatically set to On.
Page 74
Fig. 40: Turning on post-column thermostatting