Column Identification System (Column Id); Pre-Column Heater - Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual

Column compartments
Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

UltiMate 3000 Series:
TCC-3000SD and TCC-3000RS Column Compartments

2.9 Column Identification System (Column ID)

The TCC is fitted with an electronic column identification system (column ID) that lets
you store column-specific information on a column-ID chip card. This information
provides a GLP-compliant overview of the column status. The chip card is connected to
the column for the column life cycle. Column identification is supported for four columns
simultaneously.
The information on the column ID chip card is continuously updated and can be reviewed at
any time in Chromeleon, as long as the column and chip card are installed in the TCC. For
more information about the column ID, see page 75.

2.10 Pre-Column Heater

The pre-column heater brings the solvent and sample to the column temperature before
they enter the column. This avoids temperature gradients in the column.
The following pre-column heaters are available for the TCC:
Description
1 µL pre-column heater:
SST, with Viper connectors
2 µL pre-column heater:
2
SST
, with Viper connectors
3
Biocompatible
7 µL pre-column heater:
SST, with Viper connectors
3
Biocompatible
11 µL pre-column heater:
SST, with Viper connectors
3
Biocompatible
1
These pre-column heaters can be used with pressures up to 125 MPa.
2
Included in the accessories kit of the TCC-3000RS
3
These pre-column heaters can be used within the pH range 2-10 and with pressures up to 34 MPa.
Page 24
Fig. 7: Pre-column heater, here with Viper connectors
1
1
1
1
Part No.
6722.0510
6722.0530
6723.0232
6722.0540
6723.0242
6722.0550
6723.0252
Operating Instructions