Repair Guide

Reassembly Instructions

Instructions
The reassembly process is simply the reverse of the
disassembly process. However, there are various points to be
aware of.
– USB/LAN connectors must rest on top of the rear panels'
EMC spring fingers.
– The main board connector from the PSU must be pushed
firmly to fully engage it.
– Take care not to trap any cables when fitting the top
clamshell.
– Analog recorder output connections
that recorder 1 is connected to J31 on the main board.
Where applicable, recorder 2 is connected to J30 on the
main board.