Pd1500A Double-Pulse Test System Calibration - Keysight PD1500A System Installation Manual

Double-pulse test rack and safety enclosure
Hide thumbs Also See for PD1500A:
Table of Contents

Advertisement

Double-Pulse Test System
Installation Guide
6
PD1500A Double-Pulse Test
System Calibration
In this chapter:
There are several calibration-related procedures that should be performed
regularly. These procedures should be performed at least daily and anytime the
ambient temperature changes ±5 °C, anytime the oscilloscope probes, or
anything in the test system are changed.
- Oscilloscope Probe Compensation
- DPT System Calibration
- Oscilloscope Probe Deskew
- Coaxial Shunt Resistor Characterization & De-embedding. Characterization
should be performed annually. A De-embedding file will be available for
Keysight InfoLine and should be installed in the oscilloscope immediately
after characterization. See "Step 4: Obtain and Install Coaxial Shunt
Resistor De-Embedding File" on page 54.
Additionally, a Full Calibration (also known as Service Calibration) is required
annually for the DSOS104 oscilloscope, 33512B Waveform Generator, B2902A
Precision Source/Measure Unit, and Heinzinger EVO Power Supply test
instruments.