System Installation

PNA, Controller, and Test Head Module Interconnections

PNA, Controller, and Test Head Module Interconnections
IMPORTANT!
The N5292A-xx4 interconnect kits contain 2.4 mm to 3.5 mm cables. It is
important to connect the 2.4 mm end to the PNA and to connect the 3.5
mm end to the N5292A test set controller. Improper connections will result
in damage to your instruments. Refer to
It is important to use two wrenches when connecting the semirigid cables
to the 1/2/3/4 port 1.85 mm outputs front ports of the PNA to avoid
stressing the connectors and causing damage to your PNA (i.e., the
2.4 mm cable end is compatible with the 1.85 mm connector of the PNA).
If not already done, open the interconnect kit and validate its contents against
Table 2-7 on page
Figure 3-1
PNA RF Port Connector to N5292A Test Set Lo Band Connector (2.4 mm)

System Setup

Mount the analyzer on top of the test set before connecting any cables to the
PNA or to the test set. The cables can then be connected after the analyzer
and test set are securely attached.
The network analyzer is heavy. It is recommended that two
individuals, or a mechanical lift be used to lift or transport the
instrument.
Keysight Self-Configured System Installation Guide
2-16.
Figure
3-1.
3-7