Replacing Capillaries on a Standard Flow Cell

When
If the capillary is blocked
Tools required
Description
Wrench, 1/4 inch
for capillary connections
Wrench, 4 mm
(for capillary connections)
Screwdriver, Pozidriv #1 PT3
Parts required
Description
For parts see
Preparations
Turn the lamp(s) off.
Remove the front cover.
Remove the flow cell, see
All descriptions in this procedure are based on the default orientation of the cell (as it is
N O T E
manufactured). The heat exchanger/capillary and the cell body can be fixed mirror
symmetrically to have both capillaries routed to the bottom or to the top (depending on the
routing of the capillaries to the column).
The fittings at the flow cell body are special types for low dead volumes and not
N O T E
compatible with other fittings.
When retightening the fittings, make sure that they are carefully tightened (handtight plus
1/4 turn with a wrench). Otherwise damage of the flow cell body or blockage may result.
1220 Infinity II LC System User Manual
"Standard Flow Cell"
on page 352.
"Remove and Install a Flow Cell"
Maintenance and Repair
Diode Array Detector (DAD)
on page 307.