11
Maintenance and Repair
Diode Array Detector (DAD)
7
Insert the Vis-lamp (flat side to the right).
9
Fix the screws.
11
Close the lower front cover.
306
8
Reconnect the lamp connector as required.
10
Place the lamp cables in the lamp cover.
Next Steps:
12
Reset the lamp counter as described in the user
interface documentation (lamps with I.D. tag cannot be
reset).
13
Turn the lamp on and give the lamp 10 minutes to warm
up.
14
Perform a
"Wavelength Verification and
Recalibration"
on page 188or a
page 180 to check the correct positioning of the
UV-lamp.
15
Perform an
"Intensity Test"
1220 Infinity II LC System User Manual
"Holmium Oxide Test"
on page 177.