Remove the Restriction Capillary
When
For column installation
Tools required
p/n
8710-0510
1
Open both front covers.
3
Unscrew the restriction capillary from the connection of
the capillary coming from sampler or heater.
1220 Infinity II LC System User Manual
Description
Wrench open 1/4 — 5/16 inch
2
Unscrew the restriction capillary from the flow cell
capillary connection. .
4
Remove the restriction capillary from the column
compartment.
Installation
Remove the Restriction Capillary
3