1220 Infinity II LC System

Installation Check

The Installation Check switches on all available modules, purges the system for
five minutes at 1 mL/min, tests the flow path by applying a pressure up to
200 bar and switches on the oven (if available) and detector.
The pump and detector are mandatory for this check; the oven and autosampler are
N O T E
optional.
The Installation Check passes if the following conditions are met:
• All modules switch on successfully within the timeout period (120 s).
• The pump achieves 200 bar after 5 min.
• The oven reaches 2 K above its actual temperature.
• The detector lamp ignites and the detector reaches a Ready state.
1220 Infinity II LC System User Manual
Test Functions and Calibration

1220 Infinity II LC System

9
143