Wavelength Recalibration Lost

Error ID: 1037
The calibration information needed for your detector to operate correctly has
been lost.
During calibration of the detector the calibration values are stored in ROM. If
no data is available in the spectrometer ROM, the error message is generated.
Probable cause
1
2

DSP Not Running

This error message comes up when the communication between the optical
unit and the main board has a problem.
Probable cause
1
2
3

No Run Data Available In Device

In a very rare case the capacity of the CompactFlash Card is not sufficient.
This could happen for example when the interrupt of LAN communication
1220 Infinity II LC System User Manual
The detector is new.
The detector has been repaired.
Random communication error.
Defective detector main board.
Defective PDA/optical unit.
Error Information
DAD Detector Error Messages
Suggested actions
Recalibrate the detector.
Please contact your Agilent service
representative.
Suggested actions
• Switch the detector off and on again at the
power switch. If the error reoccurs:
• Please contact your Agilent service
representative.
Please contact your Agilent service
representative.