Detector Description
Match the Flow Cell to the Column
Correction factors for DAD flow cells
Table 25
Correction factors for flow cells
Flow cell
Standard flow cell, 10 mm, 13 µL, 120 bar ( 12 MPa)
(G1315-60022)
Semi-micro flow cell, 6 mm, 5 µL, 120 bar ( 12 MPa)
(G1315-60025)
Micro flow cell, 3 mm, 2 µL, 120 bar ( 12 MPa) (G1315-60024)
Semi-nano flow cell kit, 10 mm, 500 nL, 5 MPa (G1315-68724)
Nano flow cell kit, 6 mm, 80 nL, 5 MPa ( G1315-68716)
Standard flow cell bio-inert, 10 mm, 13 µL, 120 bar ( 12 MPa)
for MWD/DAD, includes Capillary Kit Flow Cells BIO (p/n
G5615-68755) (G5615-60022)
Be aware that there are additional tolerance of gasket thickness and its compression ratio,
N O T E
which are considered to be very small in comparison with the machining tolerance.
140
Path length
Correction factor
(actual)
9.80±0.07 mm
10/9.8
5.80±0.07 mm
6/5.8
3.00+0.05 mm/-0.07
3/3
mm
10.00±0.02 mm
10/10
6.00±0.02 mm
6/6
9.80±0.07 mm
10/9.8
1220 Infinity II LC System User Manual