11
Maintenance and Repair
Diode Array Detector (DAD)
Maintenance of the Flow Cell
When
If the flow cell needs repair due to leaks or contaminations (reduced light throughput)
Tools required
Description
Wrench, 1/4 inch
for capillary connections
Hexagonal key, 4 mm
Toothpick
Parts required
Description
For parts, see
Preparations
Turn the flow off.
Remove the front cover.
Remove the flow cell, see
310
"Standard Flow Cell"
on page 352.
"Remove and Install a Flow Cell"
on page 307.
1220 Infinity II LC System User Manual