Error Information
General Error Messages
Leak Sensor Short
Error ID: 0082
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak sensor current to
change within defined limits. If the current increases above the upper limit,
the error message is generated.
Probable cause
1
2
Remote Timeout
Error ID: 0070
A not-ready condition is still present on the remote input. When an analysis is
started, the system expects all not-ready conditions (for example, a not-ready
condition during detector balance) to switch to run conditions within one
minute of starting the analysis. If a not-ready condition is still present on the
remote line after one minute the error message is generated.
Probable cause
1
2
3
200
Defective leak sensor.
Leak sensor incorrectly routed, being
pinched by a metal component.
Not-ready condition in one of the
instruments connected to the remote line.
Defective remote cable.
Defective components in the instrument
showing the not-ready condition.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Ensure the instrument showing the not-ready
condition is installed correctly, and is set up
correctly for analysis.
Exchange the remote cable.
Check the instrument for defects (refer to the
instrument's documentation).