Agilent Technologies 1220 Infinity II LC User Manual page 138

Infinitylab lc series
Hide thumbs Also See for 1220 Infinity II LC:
Table of Contents

Advertisement

8
Detector Description
Match the Flow Cell to the Column
Therefore, flow cells with longer path lengths yield higher signals. Although
noise usually increases a little with increasing path length, there is a gain in
signal-to-noise ratio. For example, the noise increases by less than 10 %, but a
70 % increase in signal intensity is achieved by increasing the path length from
6 mm to 10 mm.
When increasing the path length, the cell volume usually increases — in our
example, from 5 μL to 13 μL. Typically, this causes more peak dispersion. As
Figure 50
gradient separation in our example.
As a rule-of-thumb, the flow cell volume should be about 1/3 of the peak
volume at half height. To determine the volume of your peaks, take the peak
width as reported in the integration results, multiply it by the flow rate and
divide it by 3.
Figure 50
Traditionally, LC analysis with UV detectors is based on comparing
measurements with internal or external standards. To check photometric
accuracy of the detector, it is necessary to have more precise information on
path lengths of the flow cells.
The correct response is:
expected response * correction factor
Details of the flow cells are shown in
page 140.
138
on page 138 demonstrates, this does not affect the resolution in the
Influence of cell path length on signal height
Table 24
on page 139 and
1220 Infinity II LC System User Manual
Table 25
on

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents