10
Error Information
General Error Messages
Probable cause
1
2
3
Timeout
Error ID: 0062
The timeout threshold was exceeded.
Probable cause
1
2
202
CAN cable disconnected.
Defective CAN cable.
Defective main board in another module.
The analysis was completed successfully,
and the timeout function switched off the
module as requested.
A not-ready condition was present during a
sequence or multiple-injection run for a
period longer than the timeout threshold.
Suggested actions
• Ensure all the CAN cables are connected
correctly.
• Ensure all CAN cables are installed
correctly.
Exchange the CAN cable.
Switch off the system. Restart the system, and
determine which module or modules are not
recognized by the system.
Suggested actions
Check the logbook for the occurrence and
source of a not-ready condition. Restart the
analysis where required.
Check the logbook for the occurrence and
source of a not-ready condition. Restart the
analysis where required.