Agilent Technologies 1220 Infinity II LC User Manual page 297

Infinitylab lc series
Hide thumbs Also See for 1220 Infinity II LC:
Table of Contents

Advertisement

Figure 65
1 Press the release buttons and remove the lower front cover to gain access to
2 Disconnect the inlet and outlet capillaries.
3 Unscrew the thumb screws in parallel and remove the flow cell.
4 Disassembling the flow cell.
1220 Infinity II LC System User Manual
Standard Flow Cell 10 mm / 14 µL
the flow cell area
a Unscrew the cell screw using a 4 mm hexagonal wrench.
b Remove the SST rings using a pair of tweezers.
c Use adhesive tape to remove the PEEK ring, the window and the gasket.
d Repeat step 1 on page 297 through step 3 on page 297 for the other
window. Keep the parts separate, otherwise they could get mixed up.
Maintenance and Repair
Variable Wavelength Detector (VWD)
11
297