Maintenance and Repair
Variable Wavelength Detector (VWD)
5 Cleaning the flow cell parts
6 Reassembling the flow cell
Always use new gaskets.
N O T E
7 Repeat the procedure for the other side of the cell.
8 Reconnect the inlet and outlet capillaries.
9 Test the flow cell for leaks. If there are no leaks, insert the flow cell into the
10 Perform Wavelength Calibration to check the correct positioning of the flow
11 Replace the lower front cover.
298
a Pour isopropanol into the cell hole and wipe clean with a piece of
lint-free cloth.
b Clean the windows with ethanol or methanol. Dry with a piece of
lint-free cloth.
a Hold the flow cell cassette horizontal and place the gasket in position.
Ensure that both cell holes can be seen through the holes of gasket.
b Place the the window on gasket.
c Place the PEEK ring on the window.
d Insert the conical springs. Make sure that the conical springs point
towards the window, otherwise the window might break when the cell
screw is tightened.
e Screw the cell screw into the flow cell and tighten it.
detector.
cell (
"Wavelength Verification/Calibration"
on page 168).
1220 Infinity II LC System User Manual