Controlling Collision Cell Flow - Agilent Technologies 7200 Operation Manual

Hide thumbs Also See for 7200:
Table of Contents

Advertisement

Controlling Collision Cell Flow

Operation Manual
The MS can be used to measure actual column flow. You inject a
small amount of air or other unretained chemical and time how
long it takes to reach the MS. With this time measurement, you
can calculate the column flow. See
page 58.
The collision cell gas flow rate is controlled by an EPC module
located in the GC. The collision cell gas flow consists of
nitrogen. The gas pressure at the EPC outlet controls the flow of
the gas. This pressure is controlled by the MassHunter Data
Acquisition Workstation or directly at the GC keypad. See
Set the Collision Cell Gas Flow Rate"
Operating in Electron Impact (EI) Mode

Controlling Collision Cell Flow

"To Calibrate the Column"
on page 61.