Agilent Technologies 1220 Infinity II LC User Manual page 279

Infinitylab lc series
Hide thumbs Also See for 1220 Infinity II LC:
Table of Contents

Advertisement

3
Insert the new needle-seat assembly. Press the seat
firmly into position.
5
Use Down to position the needle approximately 2 mm
above the seat.
Next Steps:
7
On completion of this procedure: Install the front cover.
8
Select End in the Tools function Change Needle.
1220 Infinity II LC System User Manual
4
Connect the seat-capillary fitting to port 5 of the
injection valve.
6
Ensure the needle is aligned with the seat. If required,
bend the needle slightly until the needle is aligned
correctly.
Maintenance and Repair
Autosampler
11
279