Maintenance and Repair
Variable Wavelength Detector (VWD)

Replace the Flow Cell / Cuvette Holder

When
If application needs a different type of flow cell or the flow cell needs repair.
Tools required
Description
Wrench, 1/4 inch
for capillary connections
Parts required
#
1
Preparations
Turn the lamp OFF.
1
Remove the lower front cover.
3
Unscrew the two thumb screws.
294
Description
Flow cell
2
Disconnect the inlet capillary and outlet tubing from the
flow cell.
4
Pull the flow cell out of its location.
1220 Infinity II LC System User Manual