Using the Cuvette Holder

This cuvette holder can be placed instead of a flow cell in the variable
wavelength detector. Standard cuvettes with standards in it, for example,
National Institute of Standards & Technology (NIST) holmium oxide solution
standard, can be fixed in it.
This can be used for wavelength verifications.
When
If your own standard should be used to checkout the instrument.
Tools required
Description
None
Parts required
p/n
G1314-60200
Preparations
Remove the normal flow cell.
Have cuvette with standard available.
1
Locate the cuvette holder on the desk.
1220 Infinity II LC System User Manual
Description
Cuvette Holder
Cuvette with the "standard", e.g. NIST certified holmium oxide sample
2
Unscrew the bracket.
Maintenance and Repair
Variable Wavelength Detector (VWD)
11
299