13
Upgrading the 1220 Infinity II LC
Oven Upgrade
Oven Upgrade
Parts required
p/n
G4297A
Software required
Lab Advisor Software
Preparations
Switch off the instrument.
1
Remove the lower front cover.
3
Remove the leak drain tube.
356
Description
1220 Infinity Oven Kit
2
Disconnect the column and remove it
Column"
4
Press the knurled portion at either side of the column
tray inwards (1) and remove the column tray (2).
(1)
Leak drain
tube
on page 55).
1220 Infinity II LC System User Manual
("Remove a
(1)
(2)