Needle Up Failed

Error ID: 4017
The needle arm failed to move successfully from the seat or out of the vial to
the upper position.
The upper position of the needle arm is monitored by a position sensor on the
sampling unit flex board. The sensor detects the successful completion of the
needle movement to the upper position. If the needle fails to reach the end
point, or if the sensor fails to recognize the needle arm movement, the error
message is generated.
Probable cause
1
2
222
Needle installed incorrectly, or wrong
needle type (too long).
Defective or dirty position sensor.
Defective motor.
Sticking spindle assembly.
Defective main board.
Defective or dirty position sensor.
Defective motor.
Suggested actions
Ensure the correct needle type is used, and
installed correctly.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.