Cell Test - Agilent Technologies 1220 Infinity II LC User Manual

Infinitylab lc series
Hide thumbs Also See for 1220 Infinity II LC:
Table of Contents

Advertisement

9
Test Functions and Calibration
Diode Array Detector (DAD)
Cell Test Evaluation
The Agilent ChemStation calculates the intensity ratio automatically. The
intensity ratio (typically between 0.5 and 0.7 for new standard flow cells and
0.1 to 0.3 for new mico- and high pressure cells) is dependent on the degree of
contamination of the flow cell windows, and on the type of flow cell used.
Figure 57
This test can be used for the standard flow cells only. The nano flow cells will give very
N O T E
low values due to their design.
Test Failed (low ratio value)
Probable cause
1
2
184
Cell Test Results (report) with no flow cell inserted
Absorbing solvent or air bubble in flow cell.
Dirty or contaminated flow cell.
Suggested actions
Ensure the flow cell is filled with water, and
free from air bubbles.
Exchange the flow cell windows.
1220 Infinity II LC System User Manual