Shutdown

Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates the
error message.
Probable cause
1
2
3
4

Lost CAN Partner

Error ID: 0071
During an analysis, the internal synchronization or communication between
one or more of the modules in the system has failed.
The system processors continually monitor the system configuration. If one or
more of the modules is no longer recognized as being connected to the system,
the error message is generated.
1220 Infinity II LC System User Manual
Leak detected in another module with a
CAN connection to the system.
Leak detected in an external instrument
with a remote connection to the system.
Shut-down in an external instrument with a
remote connection to the system.
The degasser failed to generate sufficient
vacuum for solvent degassing.
Error Information
General Error Messages
Suggested actions
Fix the leak in the external instrument before
restarting the module.
Fix the leak in the external instrument before
restarting the module.
Check external instruments for a shut-down
condition.
Check the vacuum degasser for an error
condition. Refer to the Service Manual for the
degasser or the pump that has the degasser
built-in.
10