vial positions on the tray. When the correction calculation is complete, the
values for both X and theta are rounded to one decimal place. On completion
of the alignment procedure, the corrected gripper positions are stored in the
module's firmware.
To ensure correct operation of the autosampler, the alignment procedure must be carried
N O T E
out in the correct sequence and in full (that is, without skipping any part).
ALS Alignment Controls
Button
Arm Up
Arm Down
Open Gripper
Close Gripper
Start >>
Continue >>
Restart
1220 Infinity II LC System User Manual
Test Functions and Calibration
Description
Rotate the gripper by
increasing theta.
Move the gripper horizontally
to the left.
Move the gripper horizontally
to the right.
Rotate the gripper by
decreasing theta.
Lifts the gripper arm.
Lowers the gripper arm.
Opens the gripper.
Closes the gripper.
Starts the execution of the
procedure. Shown only at the
start.
Jumps to the next step of the
procedure. Shown only during
alignment.
Restarts the execution of the
step.
Autosampler
Keyboard Shortcut
Cursor Up
Cursor Left
Cursor Right
Cursor Down
Page Up
Page Down
Enter
Enter