Illegal Temperature Value from Sensor on Main Board
Error ID: 1071
This temperature sensor (located on the detector main board) delivered a
value outside the allowed range. The parameter of this event equals the
measured temperature in 1/100 centigrade. As a result the temperature
control is switched off.
Probable cause
1
2
Illegal Temperature Value from Sensor at Air Inlet
Error ID: 1072
This temperature sensor delivered a value outside the allowed range. The
parameter of this event equals the measured temperature in 1/100 centigrade.
As a result the temperature control is switched off.
Probable cause
1
2
UV Lamp Current
Error ID: 7450
The UV lamp current is missing.
1220 Infinity II LC System User Manual
Defective sensor or main board.
Detector is exposed to illegal ambient
conditions.
The temperature sensor is defect.
Detector is exposed to illegal ambient
conditions.
Error Information
General Detector Error Messages
Suggested actions
Please contact your Agilent service
representative.
Verify that the ambient conditions are within
the allowed range.
Suggested actions
• Replace the cable to the main board.
• Please contact your Agilent service
representative.
Verify that the ambient conditions are within
the allowed range.