Maintenance and Repair
Diode Array Detector (DAD)
Next Steps:
3
Using a 4-mm hex key, tighten the window screw hand tight plus a quarter turn.
4
Reconnect the capillaries, see
5
Perform a leak test.
6
Insert the flow cell.
7
Replace the front cover
8
Perform a
"Wavelength Verification and Recalibration"
the correct positioning of the flow cell.
312
"Remove and Install a Flow Cell"
on page 188 or a
on page 307.
"Holmium Oxide Test"
1220 Infinity II LC System User Manual
on page 180 to check