11
Maintenance and Repair
Diode Array Detector (DAD)
Replacing Leak Handling System Parts
When
If the parts are corroded or broken
Tools required
None
Parts required
#
1
1
1
Preparations
Remove the front cover.
1 Pull the leak funnel out of the leak funnel holder.
2 Pull out the leak funnel with the tubing.
3 Insert the leak funnel with the tubing in its position.
4 Insert the leak funnel into the leak funnel holder.
5 Replace the front cover.
322
p/n
5041-8388
5041-8389
5062-2463
Figure 67
Replacing Leak Handling System Parts
Description
Leak funnel
Leak funnel holder
Corrugated tubing, PP, 6.5 mm id, 5 m
1220 Infinity II LC System User Manual